Application of lactate amperometric sol-gel biosensor to sequential injection determination of l-lactate

Sofia Piedade Gomes, Martina Odložilíková, M. Gabriela Almeida, Alberto N. Araújo, Cristina M.C.M. Couto, M. Conceição B.S.M. Montenegro

Research output: Contribution to journalArticlepeer-review

24 Citations (Scopus)

Abstract

This work describes the construction and evaluation of lactate sol-gel biosensors to accomplish the determination of lactate in pharmaceutical products. Lactate oxidase was incorporated in a porous sol-gel film placed onto a platinum-based electrode. Acid and basic catalysis were assessed. When coupled to a sequential injection system (SIA) the biosensor, based on (3-aminopropyl)trimethoxysilane, 2-(3,4-epoxycyclohexyl)ethyl-trimethoxysilane, deionised water, polyethylene glycol 6000 and acid catalyst, presented a range of linearity of 5 × 10-5 to 5 × 10-3 M. The analytical usefulness of the developed biosensor was evaluated through analysis of commercial pharmaceutical products containing lactate with a sampling rate of 40 samples h-1. The enzyme remained active for at least 30 days, enabling about 700 determinations without sensitivity decrease.

Original languageEnglish
Pages (from-to)1376-1381
Number of pages6
JournalJournal of Pharmaceutical and Biomedical Analysis
Volume43
Issue number4
DOIs
Publication statusPublished - 12 Mar 2007
Externally publishedYes

Keywords

  • LOx amperometric biosensor
  • Pharmaceutical products
  • SIA
  • Sol-gel

Fingerprint

Dive into the research topics of 'Application of lactate amperometric sol-gel biosensor to sequential injection determination of l-lactate'. Together they form a unique fingerprint.