Avançar para navegação principal Avançar para pesquisar Avançar para conteúdo principal

Validation of a high-performance liquid chromatographic method for the determination of doxycycline in turkey plasma

  • M. D.F. Santos
  • , H. Vermeersch
  • , J. P. Remon
  • , M. Schelkens
  • , P. De Backer
  • , R. Ducatelle
  • , F. Haesebrouck

Resultado de pesquisa: ???type-name??????researchoutput.researchoutputtypes.contributiontojournal.article???revisão de pares

49 Citações (Scopus)

Resumo

A high-performance liquid chromatographic method for the analysis of doxycycline in turkey plasma samples using demeclocycline hydrochloride as the internal standard was developed, optimized and validated. A one-step extraction procedure and an isocratic HPLC method with UV detection were used. No interferences with endogenous compounds or with the anticoagulant were observed. Linear calibration curves (r2>0.99) were obtained in water and plasma between 0 and 600 μg ml-1. Good recoveries for doxycycline (>66%) and demeclocycline (>72%) were seen both in water and in plasma. The coefficient of variation was <9.86% for within-day reproducibility and <7.53% for the between-day reproducibility. The deviation between the mean value found and the true value was <14.5% (accuracy). The limit of detection was 0.1 μg ml-1 in plasma samples. A good stability of doxycycline was observed in water and in plasma samples after storage for six months at -20°C (recovery >91%).

Idioma original???core.languages.en_GB???
Páginas (de-até)301-308
Número de páginas8
RevistaJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
Volume682
Número de emissão2
DOIs
Estado da publicação???researchoutput.status.published??? - 12 jul. 1996
Publicado externamenteSim

Impressão digital

Mergulhe nos tópicos de investigação de “Validation of a high-performance liquid chromatographic method for the determination of doxycycline in turkey plasma“. Em conjunto formam uma impressão digital única.